LUPIN BF3-NP中子周围剂量率仪
NEUTRON REM COUNTER FOR PULSED FIELDS
LUPIN BF3-NP is an environmental monitoring unit for neutron H*(10) measurements, with unique excellent performance for neutron detection in pulsed fields.
The rem counter is composed of:
BF3 neutron proportional counter
Cylindrical moderating assembly
Built-in power supply, signal acquisition and processing, and control electronics
The built-in electronics processes the signal coming from the detector and elaborates the instantaneous H*(10) rate value every second.
If required, the radiation sensitive electronics can be housed in a separate rack, thanks to the modularity of the instrument components.
The data are sent to the connected SATURN ratemeter acquisition and control unit, which locally displays the instantaneous H*(10) rate and the integrated values and compares them to the pre-set alarm thresholds.
A LUPIN BF3-NP LITE version is also available, featuring a lighter build and a narrower energy range, suitable for all the applications that do not require to detect neutrons with energy above 20 MeV.
Main features:
Only detector world-wide efficiently working in pulsed fields
Modular electronics
High sensitivity
Fully customizable for specific requirements of particle accelerator facilities
Energy range: from thermal up to 10 GeV or up to 20 MeV (LITE version)
Energy response closely resembles the ICRP74 conversion curve
Unaffected by signal pile-up and so particularly suited for pulsed neutron fields
Maximum H*(10) per burst in pulsed fields: 2 µSv
Connectable to SATURN ratemeters
Excellent gamma rejection (< 0.5 µSv/h at 50 mSv/h, 662 keV)