EJ-276 n、γ 甄别探测器

产品详情



EJ-276 pulse-shape discriminating (PSD) plastic scintillator enables the separation of gamma and fast neutron signals on the basis of their timing characteristics. This scintillator replaces all versions of EJ-299-33 and EJ-299-34 PSD scintillators and embodies the following improvements:

  • Excellent physical hardness, equal to or superior to that of standard plastic scintillators

  • Long-term stability of scintillation and optical characteristics

  • Basic PSD properties increased to being comparable to the best liquid scintillators

EJ-276G with green fluorescence is also available for use with solid-state sensors.



 PROPERTIES

EJ-276

EJ-276G

Light Output (% Anthracene)

56

52

Scintillation Efficiency   (photons/1 MeV e-)

8,600

8,000

Wavelength of Maximum Emission   (nm)

425

490

No. of H Atoms per cm3 (x1022)

4.546

4.546

No. of C Atoms per cm3 (x1022)

4.906

4.906

No. of Electrons per cm3 (x1023)

3.533

3.533

Density (g/cm3)

1.096

1.096

Approx. Mean Decay Times of   First 3 Components (ns) 

Gamma Excitation

13, 35, 270

Neutron Excitation

13, 59, 460





Chemical Compatibility

Attacked by:  Aromatic solvents, Chlorinated solvents, Ketones, Solvent bonding cements, etc.

Stable in:  Water, Dilute acids and alkalis, Lower alcohols, and Silicone greases.

It is safe to use most epoxies with these scintillators.


EJ-276_emspec.png